![Argus 707 V Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/argus/707-v/707-v_service-manual_2969492043.webp)
REPLACEMENT OF PARTS
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
42 / 48
7. Use the short screw at these
places (torque 1.1 Nm)!
7. Use the long screw at these
places (torque 1.1 Nm)!
Mind the torque for the screws listed below:
What:
Where
Torque in Nm
Case Backplane
into
front
1.10
Pump unit
Into front
1.00
Combination clamp
Into clamp holder
1.50
Pressure sensor
Into front
0.60
Stop flow
Into front
0.60
Air detector
Into front
0.60
Sensor board
Into pump unit
0.40
Display board
Into front
0.50
Transformer Into
Backplane
1.00
Clamp holder
Into Backplane
0.50
Mains plug
Into Backplane
0.50
Edge board
Into clamp holder
0.50