SOFTWARE UPDATES
14.203.B_A707 en.SM.V4.3X
ARGUS Medical AG 28.03.06 / PJ
27 / 48
7. Select the requested pump firmware file by
pressing the button “…”.
8. Press “Update the pump firmware”. Follow the
instructions displayed on the PC. The firm-
ware will be installed and the pump will be
switched off automatically.
9. Go into the configuration part again (refer to point 2-3).
Press the “Edit” button and enter the pump PIN code (de-
fault PIN after firmware update is 0).
10. Restore the old configuration (incl. calibration) from the
previous
created file.
11. Restore the configuration by pressing the “Download” but-
ton.
12. Perform a standard safety check
(see chapter 10)
, nor-
mally the calibration will not be destroyed if the procedure
is carefully performed step by step
.
5.3.3. Upgrade of a pump with firmware < V4.xx
With the “Update center” it is also possible to upgrade pump firmware older than
V4.xx.
Important remark!
The actual calibration (and configuration) will be stored temporary on the PC,
please perform the upgrade procedure pump by pump. Otherwise invalid cali-
bration values will be stored on the pump.
It is urgent necessary to perform a standard safety check (
see chapter 10
)!
1. Go into the “Update center” (s
ee point 1-4
of chapter 5.3.2.
):
2. Switch the pump ON by keeping the key
[10] pressed.
3. Backup the legacy configuration (present
configuration before the firmware update).
This may take several seconds.
4. Switch the pump OFF.
5. Press “Upgrade to bootloader 4”. Follow
the instructions displayed on the PC. The
bootloader will be upgraded then.
6. Select the requested pump firmware file
by pressing the button “…”.
7. Press “Update the pump firmware”. Follow
the instructions displayed on the PC. The
firmware will be installed and the pump
will be switched off automatically.