CHAPTER 4
REMOTE CONTROL PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 72 of 93
44M00
[STATe:] NG? {; | NL}
Command Syntax:
[STATe:] NG? {; | NL}
Purpose:
Query if the NG flag is displayed on this module.
Description:
This command queries the NG status. If the response is “0”, the LED of NG (NO GOOD)
will be off. If the response is ”1”, the LED will be lit, showing that the NG flag is present.
[STATe:] PROTect? {; | NL}
Command Syntax:
[STATe:] PROTect? {; | NL}
Purpose:
Query the state of the protection register on this module.
Description:
1.
PROT? requests the status of the units protection register.
2.
Use the command “CLR” to clear the register of PROT status to “0”.
BIT ID
BIT VALUE
REMARK
bit 0
0 = Off, 1 = Triggered
Over Power Protection (OPP)
bit 1
0 = Off, 1 = Triggered
Over Temperature Protection (OTP)
bit 2
0 = Off, 1 = Triggered
Over Voltage Protection (OVP)
bit 3
0 = Off, 1 = Triggered
Over Current Protection (OCP)
Table 23: Protection Status Register
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit1 Bit 0
7
6
5
4
3
2
1
0
Over Power Protection (OPP)
Over Voltage Protection (OVP
)
Over Current Protection (OCP
)
Over Temperature Protection (OTP)