CHAPTER 4
REMOTE CONTROL PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 58 of 93
44M00
1. The least significant number is the fifth digit after the decimal point.
2. The LOW level current value cannot be higher than the HIGH level.
3. Should a value be entered that is higher than what is possible then the 44M04 will
automatically set the maximum value according the load module installed.
4. The engineering unit is A.
CURR
Command Syntax:
[PRESet
:
] CC | CURR {SP} {NR2}{: | NL}
[PRESet
:
] CC | CURR?{
; |
NL}
Purpose:
Set or read the load current.
Description:
This command is used to set or query the load current.
1. The least significant number is the fifth digit after the decimal point.
2. Should a value be entered that is higher than what is possible then the 44M04 will
automatically set the maximum value according the load module installed.
3. The engineering unit is A.
CP
: {
HIGH | LOW}
Command Syntax:
[PRESet
:
] CP
:
{ HIGH | LOW} {SP} { NR2} {
; |
NL}
[PRESet
:
] CP
:
{ HIGH | LOW}? {
;
| NL}
Purpose:
Set and read the operating power value in Watts
Description:
This command is used to set or query the HIGH or LOW setting levels of load power.
These 2 power levels need to be used should a dynamic load waveform be desired. It also
allows the user to switch between two preset power levels.
1. The least significant number is the fifth digit after the decimal point.
2. The LOW level power value cannot be higher than the HIGH level.
3. Should a value be entered that is higher than what is possible then the 44M04 will
automatically set its maximum value according the load module installed.
4. The engineering unit is W.
CR | RES
:
{HIGH | LOW}
Command Syntax:
[PRESet
:
] CR | RES
:
{HIGH | LOW} {SP} {NR2} {
;
| NL}
[PRESet
:
] CR | RES
:
{HIGH | LOW}? {
; |
NL}