44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 68 of 93
44M00
[LIMit
:
]VOLtage
:
{HIGH | LOW} or VH | VL
Command Syntax:
[LIMit:]VOLtage
:
{HIGH | LOW} {SP} { NR2 }{
;|
NL}
[LIMit:]VOLtage
:
{ HIGH
LOW}?
{
;|
NL}
[VH | VL] {SP} { NR2 }{
;|
NL}
[VH | VL]? {
;|
NL}
Purpose:
Set or query the HIGH / LOW limits of voltage present at the load terminals.
Description:
This command is used to set two voltage LIMIT values. Operation outside these LIMIT
values will cause a NG signal to be generated.
1. The LOW level cannot be higher than the HIGH level.
2. If the voltage at the load input falls below the LOW limit then a No Good (NG) signal is
available.
3. If the voltage rises above the HIGH limit then the NG signal is available.
4. If the current stays between HIGH and LOW LIMIT levels, the NG signal will not be
sent.
[LIMit
:
] SVH
SVL
Command Syntax:
[LIMit
:
] {SVH | SVL} {SP} {NR2}{
;|
NL}
[LIMit
:
] {SVH | SVL}? {
;|
NL}
Purpose:
Set or query the upper and lower voltage levels during for short test.
Description:
This command is used to set two voltage LIMIT values. If during the short test the voltage
is outside these LIMIT values, a NG signal will be generated.
1. The LOW level cannot be higher than the HIGH level.
2. If the voltage at the load input falls below the LOW limit then a No Good (NG) signal is
available.
3. If the voltage rises above the HIGH limit then the NG signal is given.
4. If the current stays between HIGH and LOW LIMIT levels, the NG signal will not be
sent.