CHAPTER 3
MAINFRAME OPERATION
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 25 of 93
44M00
# Displays
Description
6
Status
Display
At power up, the LCD will display “Nor” which indicates NORMAL operation. The message
displayed will change if a setting such as WAKE-UP or AUTO-SEQUENCE is selected.
7
BANK T1
/ Display
The upper digits on the right hand side of the screen relate to the memory BANK in normal
mode. There are 15 BANKS which can be selected in turn by pressing the upper pair of arrow
keys. Each BANK has 10 separate memory STATES (locations) which are selected with the
lower pair of arrow keys.
When in auto-sequence mode, T1 is displayed. T1 is the test time. The test time can be adjusted
using the upper arrow keys between 0.1sec and 9.9sec in 100ms steps. Please note that during
the T1 test time, the mainframe LCD will not flag “NG”. The T2 setting is used for checking
the no-go (NG) function according to the voltage, current or power limits set in the LIMIT
menu.
8
STATE /
T2
Display
The lower digits relate to the memory STATE (location) in normal mode. There are 10
memory STATES (locations) which can be selected in turn by pressing the Lower arrow keys.
These memory STATES are supplemented by the 15 memory BANKS, giving the user a total
of 150 memory locations.
When in auto-sequence mode the T2 function is displayed. T2 is the time that NG/GO is
checked according to the LIMITS that have been set for that test step. So if the NG flag has
been enabled and the load measures values outside the preset LIMIT values, then the test will
stop during T2. The mainframe’s LCD will flash “NG” and the test will stop at that step in the
auto-sequence. At this point, the user can either press the ENTER key to carry on to the next
step or the EXIT key to abort the auto-sequence.
9
Arrow
Buttons
The ARROW keys are used to increase or decrease setting values.
Table 3: 44M04 Mainframe LCD Call-outs