Portable Test Equipment
User’s Guide
PTE-UGFW56a-EN
www.appliedp.com
88
SYST:INP:CONF {<x>}
Setting the measuring inputs combination. Parameter can have the following values:
x=0 when only internal voltage and current is being measured, regardless of connected measuring
accessory
x=1 when accessory connected to Input A is measured together with internal voltage
x=2 when accessory connected to Input B is measured together with internal current
x=3 when only accessories connected into Inputs A and B are measured, regardless of generated
voltage or current
These measured values are shown on blue screens and corresponding blue type screens are visible
for each particular combination: Load Values when U and I are measured together, U+U screens
when voltages only are available, etc.
SYST:INP:CONF?
Request for inputs configuration. Answer can be 0 ÷ 3 with the same description as for previous
(setting) command.
Results Reading Commands
RES:COUN:EMER?
Request for number of stored meters.
RES:COUN:LP? {<emer (1-255)>}
Request for number of stored load-points of defined meter.
RES:GET:EMER? {<emer (1-255)>}
Request for all header data of defined meter.
RES:GET:LP? {<emer (1-255)>},{<lp (1-255)>}
Request for all load-point data of defined meter and load-point.
RES:CLR
Clearing of all results (stored meters
’ data).
Measurement Commands
Measurement commands return measured values in form of a set of three real numbers without unit.
Values are in classic or exponential format (depends on value and firmware version) with max. 6
decimal places and separation of decimal part of number is made by point.
Three values are separated by comma or comma followed by space (depends on firmware version).
Examples of return string:
220.123456,221.234567,219.345678
+3.456789E-03, +4.567891E-03, +5.678912E-03
VOLT:AC:RANG?
Request for the voltage range in connected voltage accessory (Voltage Transducer). Answer can be
0, 1 … (up to supported number of ranges).
VOLT:AC:RANG {<range>}
Setting the voltage range in connected voltage accessory (Voltage Transducer). Parameter can be 0,
1 … (up to supported number of ranges).
VOLT:AC:RANG:AUTO?
Request for the status of auto voltage range in connected voltage accessory (Voltage Transducer).
Answer is: 0 = OFF, 1 = ON.
Содержание PTE 2100 Series
Страница 1: ...Portable Test Equipment Models PTE 2100 PTE 2300 User s Guide Version 5 6a ...
Страница 102: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 102 7 APPENDIX ...
Страница 104: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 104 Figure 22 3P3W Connection with OPTS ...
Страница 105: ...Portable Test Equipment User s Guide www appliedp com PTE UGFW56a EN 105 Figure 23 1P2W Connection with OPTS ...