Portable Test Equipment
User’s Guide
www.appliedp.com
PTE-UGFW56a-EN
31
Current Standby
.............. behavior of current channel when zero current should be generated.
Passive
means that no current is generated nor measured. However,
there can be some residual current present in current loop. When current
loop is disconnected, nothing happens.
Active
means that Generator
generates zero current which it tries also to compensate this current.
Actively generates zero current. When current loop is compromised error
in current generation is indicated.
MEASUREMENT:
Active Voltage
........ enable (
ON
) or disable (
OFF
) voltage presence between tests when
Automatic Mode is in progress
Test ident.
.............. test identification by
NAME
or
CODE
in table which is shown in
Automatic mode, see picture on page 27.
Back Control timing depends on integration time of regular measurement. User has to take into
account that Back control will last at least two times more than actual integration time (around
4s + 2 x integration time). Integration time is a modifiable parameter - see Time base parameter in
chapter 5.7.1 Measurement Settings.
4.3.2 Inputs Configuration
Inputs Configuration Screen
* P150
This
screen
allows
switching
between
several
measurement inputs.
1. Generator U and I
This is the only option here, which allows the user to
enter the basic Generator screen (like depicted in
chapter 3.2) to set all possible parameters (related of
generating power). After pressing
OK
when the
cursor is located on item no. 1 the basic generator
screen appears.
Blue screens show load values (U, I, energy, power,
meter error measurement possible, etc.)
2. Ext. [In A] & Internal U
In this option external accessory connected to
input
A
and
internal voltage
are selected to be measured
together. User cannot change generated parameters;
all changes are possible only when item no. 1 is
selected in this menu. However, Generator can be in
running state if it was started previously.
In this option only this screen is shown to the user
from all generator’s screens (green ones).
When this option is pressed main blue screen appears with regards of used accessory. If
connected accessory is for voltage measurement; U+U screen is shown, otherwise it is the Load
Values screen.
3. Internal I & Ext. [In B]
In this option external accessory connected to
input B
and
internal current
are selected to be
measured together. User cannot change generated parameters; all changes are possible only
when item no. 1 is selected in this menu. However, Generator can be in running state if it was
started previously.
In this option only this screen
is shown to the user from all generator’s screens (green ones).
When this option is pressed main blue screen appears with regards of used accessory. If
connected accessory is for current measurement; I+I screen is shown, otherwise it is the Load
Values screen.
Содержание PTE 2100 Series
Страница 1: ...Portable Test Equipment Models PTE 2100 PTE 2300 User s Guide Version 5 6a ...
Страница 102: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 102 7 APPENDIX ...
Страница 104: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 104 Figure 22 3P3W Connection with OPTS ...
Страница 105: ...Portable Test Equipment User s Guide www appliedp com PTE UGFW56a EN 105 Figure 23 1P2W Connection with OPTS ...