Load Values screens for single phase system:
There are only two versions of Load Values screens:
Load Values screen
* P44
Load Values Primary screen
* P69
Primary screen shows primary values which depend on CT and VT ratios defined for actual meter.
5.1.2 Actual Values
Actual Values
screens show actual measured and calculated values of load point when other
combination than one current and one voltage measurement sensors are used.
Device shows measured and calculated values of Current (
CT
) and Voltage (
VT
) Transformers.
In case of connection of one measuring sensor device shows:
basic quantities (depends on used sensor)
Transmission Ratio
r
Lx/L1
Transmission Error
er
Lx/L1
Phase Error
φ
Lx/L1
Left values are comparison of values in phases L2 and L1 and right values in phases L3 and L1.
In case of connection of two measurement sensors device shows:
basic quantities (depends on used sensors)
Transmission Ratio
r
B/A
Transmission Error
er
B/A
Phase Error
φ
B/A
Showed values in each phase are comparison of values in Input A and Input B.
Transmission Error is calculated as comparison of measured Transmission Ratio (
r
Lx/L1
or
r
B/A
) and
nominal Transmission Ratio (
r
nom
).
Content of
Actual Values
screens depends on connected measuring accessories.
Содержание PTE 2100 Series
Страница 1: ...Portable Test Equipment Models PTE 2100 PTE 2300 User s Guide Version 5 6a ...
Страница 102: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 102 7 APPENDIX ...
Страница 104: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 104 Figure 22 3P3W Connection with OPTS ...
Страница 105: ...Portable Test Equipment User s Guide www appliedp com PTE UGFW56a EN 105 Figure 23 1P2W Connection with OPTS ...