Portable Test Equipment
User’s Guide
www.appliedp.com
PTE-UGFW56a-EN
83
SYST:ENER:DOS:STOP
Stop (or cancel) of dosage function.
SYST:ENER:DOS?
Request for general status of dosage function. If dosage function is active reply is: percentage, done
energy or time and full energy or time.
SYST:ENER:DOS:STAT?
Request for status of dosage function. Reply is 0 (dosage not active), 1 (dosage initialized), 2 (dosage
started and running) or 3 (dosage stopped).
SYST:ENER:DOS:STAT:PERC?
Request for done percentage of dosage.
SYST:ENER:DOS:STAT:VAL?
Request for done value (time or energy amount) of dosage.
SYST:ENER:DOS:STAT:TIM?
Request for time of
“time dosage” in seconds.
SYST:ENER:DOS:STAT:ENER?
Request for energy of
“energy dosage” in Ws resp. VAs.
SYST:ENER:DOS:STAT:ENER:K?
Request for done energy of
“energy dosage” in kWh resp. kVAh.
SYST:ENER:DOS:MOD {<mode>}
Setting of Dosage Mode. Parameter is number (0 -
5) or code (TIME, ACTIVE …) of Dosage Mode.
SYST:ENER:DOS:MOD?
Request for state of Dosage Mode. Answer is number (0 -
5) or code (TIME, ACTIVE …) of Dosage
Mode.
SYST:ENER:DOS:TYP {<type>}
Setting of Dosage Type. Parameter is 0 for time or 1 for energy.
SYST:ENER:DOS:TYP?
Request for state of Dosage Type. Answer is 0 for time or 1 for energy.
SYST:ENER:DOS:TIM {<time>}
Setting of time (in seconds) for time dosage.
SYST:ENER:DOS:TIM?
Request for time for
“time dosage”. Answer is time in seconds.
SYST:ENER:DOS:ENER {<energy>}
Setting of energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER?
Request for energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER:K {<energy>}
Setting of energy (in kWh or kVAh) for energy dosage.
Содержание PTE 2100 Series
Страница 1: ...Portable Test Equipment Models PTE 2100 PTE 2300 User s Guide Version 5 6a ...
Страница 102: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 102 7 APPENDIX ...
Страница 104: ...Portable Test Equipment User s Guide PTE UGFW56a EN www appliedp com 104 Figure 22 3P3W Connection with OPTS ...
Страница 105: ...Portable Test Equipment User s Guide www appliedp com PTE UGFW56a EN 105 Figure 23 1P2W Connection with OPTS ...