Creating and Editing Methods 5-15
=
4
From the Insert screen, press
Pause
. The
Programmed Pause
screen appears.
5
In the
Start first Pause at Cycle
field, type in the cycle number where you want the
method to first pause.
6
In the
Pause Every
field, type in the pause frequency in cycles.
The pause frequency specifies the number of cycles that will run between each
pause.
7
In the
Cycles For
field, type in the length of the pause in minutes:seconds
(00:01–98:59) format.
8
In the
Beep During The Pause?
field, press
Yes
or
No
.
9
Press
Accept
to accept the pause information on the screen.
The word
Pause
now displays to the right of the incubation step where you
programmed the pause.
Note
You can only insert one pause in each cycle.
To insert a programmed pause:
(continued)
Step
Action
Define Programmed Pause
Start first pause at cycle 1 of 25.
Pause every 25 cycles for 00:30.
Beep during the pause? No
F1
F3
F4
F5
F2
Accept