3-18 Instrument Setup
Using the T
m
Calculator
How to Use the
T
m
Calculator
Use the T
m
Calculator to determine the annealing temperature of a primer set of
known sequence.
To use the T
m
Calculator:
Step
Action
1
From the
Main
menu, press
Util
. The Utilities screen opens as shown below.
2
Press
TmCalc
.
The Tm Calculator appears.
3
Enter the salt concentration.
Note
The default is 50. Enter values 5 to 1000.
4
Enter the primer concentration.
Note
The default it 0.20. Enter values 0.01 to 10.00.
5
Enter primer sequence in P1.
6
Enter primer sequence in P2 and press
Enter
to calculate the T
m
s.
The melting points are displayed. Use this information to program a run.
For more information, see Chapter 5, “Creating and Editing Methods.”
7
Press
Return
to display the
Utilities
screen.
Utilities
Diag - Instrument diagnostics
TmCalc - Calculates melting temp
Config - Instrument configuration
F1
F3
F4
F5
F2
Diag
Exit
More
Config
TmCalc
[Salt]: 50 mM
[Primer] 0.20 uM
P1: 5’
P2: 5’
Tm of P1=
Tm of P2=
Press ENTER to calculate Tm’s