<SYSTEM CONFIG>Page
unchangeable, these parameters include:
Start bits: 1-bit
Data bits: 8-bit
Stop bits: 1-bit
Verifying mode: none
5.8.3
Command Handshake
After command handshake is turned on, and each time receiving command, instrument will
return an identification code to the host after finishing process.
For example:
Transmit: DISP: PAGE SETUP
Return: E0
Table 5-1 Identification code definition return by command handshake
Identification code
Meaning
E0
Normal communication, no error.
E2
Command error
E3
Parameter error
E4
Invalid parameter
E5
Separator error
E6
Numeric parameter error
E7
Numeric Over length
E8
Invalid command