64
AT4610/AT4710 User’s Guide
11.6.7
SYST:CJC
SYST:CJC Cold junction compensation switch.
Command syntax
SYST:CJC {ON,OFF,0,1}
Query syntax
SYST:CJC?
Query response
{ON,OFF}
On <DATALOG> page, if logging has been started up, this command is invalid.
11.6.8
SYST:OCD
SYST:OCD Thermocouple open circuit checking switch.
Command syntax
SYST:OCD {ON,OFF,0,1}
Query syntax
SYST:OCD?
Query response
{ON,OFF}
On <DATALOG> page
,
if logging has been started up, this command is invalid.
11.6.9
SYST:FILTER
SYST:FILTER Digital filter switch.
Command syntax
SYST:FILTER {ON,OFF,0,1}
Query syntax
SYST:FILTER?
Query response
{ON,OFF}
On <DATALOG> page, if logging has been started up, this command is invalid.
11.6.10
SYST:RESULT:FONT
SYST:RESULT:FONT Select font for measuring value display on <DATALOG> page and
<MEAS DISPLAY> page.
Command syntax
SYST:RESULT:FONT {SYSTEM,SWISS,ENVY,CPM}
Query syntax
SYST:RESULT:FONT?
Query response
{SYSTEM,SWISS,ENVY,CPM}
11.6.11
SYST:RESULT:COLOR
SYST: RESULT:COLOR Configure color for measuring value display on <DATALOG> page
and <MEAS DISPLAY> page.
Command syntax
SYST:RESULT:COLOR {CHANNEL,WHITE}
Parameter
there
,
CHANNEL: and channel set color in <SETUP> page
WHITE: white color
Query syntax
SYST:RESULT:COLOR?
Query response
{CHANNEL,WHITE}
11.7
COMPARATOR SET SUBSYSTEM
Table 11-4
COMPARATOR subsystem tree
COMP:EN
Comparator main switch
COMP:CH:EN
Comparator switch of an appointed channel
COMP:CH
COMPARATOR SET of an appointed channel