73
Specifications and Calibration
TC-S
Table 12-8 Specifications for thermocouple TC-S
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
0 ~ 100 °C
±4.5 °C
500 °C
0.05 °C
0 ~ 100 °C
±4.5 °C
100.00 ~ 300.00 °C
±3.0 °C
300 ~ 500 °C
±2.2 °C
2000 °C
0.1 °C
0 ~ 100 °C
±4.5 °C
100.00 ~ 300.00 °C
±3.0 °C
300 ~ 1700 °C
±2.2 °C
TC-B
Table 12-9 Specifications for thermocouple TC-B
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
-
-
-
500 °C
0.05 °C
250 ~ 500 °C
±1.8 °C
2000 °C
0.1 °C
0 ~ 1800 °C
±1.8 °C
TC-R
Table 12-10 Specifications for thermocouple TC-R
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
0 ~ 100 °C
±4.5 °C
500 °C
0.05 °C
0 ~ 100 °C
±4.5 °C
100.00 ~ 300.00 °C
±3.0 °C
300 ~ 500 °C
±2.2 °C
2000 °C
0.1 °C
0 ~ 100 °C
±4.5 °C
100.00 ~ 300.00 °C
±3.0 °C
300 ~ 1700 °C
±2.2 °C
12.3.2
PT100 RTD
Figure 12-4 Connection of 3-wire RTD
Figure 12-5 Connection of 4-wire RTD
Executive standard JIS C1604-1997, IEC751, JB/T8622-1997, JB/T8623-1997
Table 12-11 PT100 RTD (RTD-3W/RTD-4W) Specifications
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100 ~ 100 °C
±0.6 °C
500 °C
0.05 °C
-200 ~ 500 °C
±0.8 °C
2000 °C
0.1 °C
-200 ~ 850 °C
±1.0 °C
12.3.3
Cu50 Copper RTD
Table 12-12 Cu50 (3W) Copper RTD Specifications