59
SCPI Command Reference
2. <Scifloat>
:
scientific notation floating number
:
1.23E+4
, +1.23e-4
3. <Mpfloat>: multiplier expressed by floating number: 1.23k, 1.23M,1.23G,1.23u
Table 11-1 Multiplier Mnemonics
Definition
Mnemonic
1E18 (EXA)
EX
1E15 (PETA)
PE
1E12 (TERA)
T
1E9 (GIGA)
G
1E6 (MEGA)
MA
1E3 (KILO)
K
1E-3 (MILLI)
M
1E-6 (MICRO)
U
1E-9 (NANO)
N
1E-12 (PICO)
P
1E-15 (PEMTO)
F
1E-18 (ATTO)
A
Multiplier is Case Insensitive, its writing style is different from standard name.
11.2.3
Separator
Instrument terminator only accepts allowed separators, terminator will occur E5 error if beyond
this separator, and these separators include:
:
colon, used for separate command tree, or restart command tree.
Example
:
AAA : BBB : CCC 123.4
?
question mark, used for query
Example
:
AAA ?
□
space, used for separate parameter
Example
:
AAA:BBB□1.234
11.3
Command Reference
All commands in this reference are fully explained and listed in the following functional
command
order, the following is all subsystem
DISPLAY
Display SUBSYSTEM Page
SETUP
Setup SUBSYSTEM
SYST
System SUBSYSTEM
COMP
Comparator SUBSYSTEM
FILE
File SUBSYSTEM
MEAS
Meas Display SUBSYSTEM
LOG
Data Logging SUBSYSTEM
Other command:
IDN?
Query instrument’s version
KEYLOCK
keypad lock
TIME
time set
ERR:SHAK
handshake command switch
11.4
DISPLAY
SUBSYSTEM
DISPlay subsystem is used for switching different display page or display a string of text on