33
<COMPARATOR SET> Page
6.2
Comparator [Upper/Lower] Limit Setting
The comparator for each channel is independently set, and there are 2 groups of independent
comparator for instrument’s each channel,
Steps of changing upper/lower limit
Step 1
Enter <COMPARATOR SET> page
Step 2
Touch screen or use cursor to select the field of one channel’s [Lower] Limit or [Upper] Limit.
Step 3
Touch function field at right side, click & select [Input Lower] or [Input Upper], input data in the popup
window:
TIP
Please don’t input lower limit value lower than upper limit value.
Lower limit
≤ measured value ≤ Upper limit
OK (green indicator
)
Measured value < Lower limit
NG (red indicator
)
Measured value > Upper limit
NG
(
red indicator
)
6.3
Display Comparative Result
Give an example of how comparator display in <MEAS DISPLAY> page:
Suppose the comparator value of all channels is as below:
Type: TC-K
Range: 100
°C
Comparator 1: turn on, range [0.00°C, 50.00°C]
Comparator 2: turn on, range [50.00°C, 100.00°C]
See as follow figure: