24 DECADE user manual, edition 9
time file when the baseline is drifting. The available drift monitor settings
depend on the range selected at time 00:00 in the time file (see page 37).
Can be set between 0.1 pA/min and 10 nA/min.
E1, E2, E3
PULS SET2
control
Controls the cell potential settings of the pulse. A change in E1 has to be
confirmed with the ‘#’-button (see page 51).
Ec
(AUTO) PRGR (DC only)
control
Controls the cell potential in a time file (see page 37).
Ecell
PULS STAT, AUTO RUN (Pulse only), SCAN STAT
status
Displays the cell potential during ‘t1’ in the pulse mode (see page 51).
Displays the actual cell potential in the scan mode (see page 59).
Eox. or Ered.
DC SET
control
Controls the oxidative or reductive cell potential. Can only be set or changed
after confirmation with the ‘#’ button. Passing the 0.00 V value automatically
changes the readout from ‘ox’ to ‘red’ and in reverse.
Eox. or Ered.
DC STAT, AUTO RUN (DC only)
status
Displays the oxidative or reductive cell potential.
File
AUTO SET
control
Controls the number of the time file to be edited or activated. In the DC mode
file numbers 1 - 5 are available, in the pulse mode file numbers 6 - 9 can be
selected. The time files remain stored in RAM, also after switching off the
DECADE (see page 37).
Filt.
DC SET, DC STAT, (AUTO) PRGR (DC ONLY)
control
Controls the rise time filter settings for the recorder output, which are: 0.1, 0.2,
0.5, 1, 2 or 5 seconds (see page 29).
HOLD=0/1
AUTO RUN, SCAN STAT
function