CHAPTER 5 AUTO mode and time files
45
9.
If ‘STOP’ is chosen from the keyboard, the execution of the time file is
interrupted. To start again, an external ‘RESET’ and ‘START’ (or ‘RUN’
from keyboard) command is required.
Software version 3.00 and higher: the STOP command toggles between
STOP and START to control the execution of a time file. STOP also
deactivates the outputs Aux 1 and 2, and Relays 1 and 2 (status: 0000).
10. When ‘RESET’ is chosen from the keyboard, the execution of the time
file is interrupted and the ‘AUTO SET’ screen appears.
Software version 3.00 and higher: RESET also deactivates the outputs
Aux 1 and 2, and Relays 1 and 2 (status: 0000).
Note: In the
master
mode the HOLD, STOP, START and RESET commands
are accessible only from the keyboard. In the
slave
mode these inputs are
accessible via the rear panel of the DECADE. Accessible keyboard
commands in the
slave
mode are: RUN, STOP, RESET and A-ZERO.
Other events
Inject/load and inject marker
A frequently used function is the inject/load (or inject marker) as a trigger for
starting the integration software. This trigger is always accessible, not only in
the AUTO mode.
Fig. 15. Output contact needed to start the integrator by switching the manual
(A) or an automated electric valve (B).
A manual injector with position sensor has an internal relay which is
connected by a cable connection in the oven cabinet of the DECADE to the
inject/load of the rear panel. The relay is open when the valve is in ‘load’
position, the relay closes when the valve is switched to ‘inject’ and opens
again when the valve is switched back to ‘load’.
The ‘INJECT MARKER’ is only active when an electrically actuated injector is
present. This output is high when the injector is in ‘load’ position and low in
the ‘inject’ position.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com