26 DECADE user manual, edition 9
‘0000’ corresponds to the four output controls at the rear panel of the
controller (from left to right: relay 2, relay 1, aux 2, aux 1). The position
indicated on rear panel refers to the ‘0000’ setting (see page 37).
PREV
DC SET, AUTO SET, PULS SET1, SCAN SET
function
Returns to previous screen.
PRGR
AUTO SET
function
Enters the (AUTO) PRGR screen
Range
DC SET, DC STAT, (AUTO) PRGR, PULS SET1, PULS
STAT, SCAN SET, SCAN STAT
control
Controls the recorder range setting, varying from 10 pA to 5 µA full scale,
divided in 1, 2 and 5 steps. In the DC mode, selectable recorder ranges
depend on the pre-selection of the integration range in the ‘Range select’
screen. In the pulse and scan mode only the µA range can be used (see
page 29).
RESET
AUTO RUN
function
Aborts the time file and returns to the ‘AUTO SET’ screen. The cycle counter
(‘Cy’) is reset to 1 (see page 37).
Software version 3.00 and higher: the outputs Aux 1 and 2, and Relays 1 and
2 are deactivated (status: 0000).
RUN
AUTO SET
function
Starts the execution of a time file (master/slave). If the DECADE is slave, the
system waits for the ‘START’ input trigger to start the next run (see page 37).
Software 3.00 and higher: In the slave mode the RUN command initialises the
AUTO RUN screen, the system waits for the ‘START’ input trigger to start a
time file.
Scan
SCAN SET
control
Controls the scan speed, can be set from 1 - 50 mV/s in 1, 2, 5 steps (see
page 59).