14 DECADE user manual, edition 9
line voltage. If not correct, this insert has to be reversed. Use only a
supply appliance with protective grounding. The fuses are included in the
line connector.
The correct values are given on the rear panel for the different line
sources.
For 110 V (AC) ± 10%, use two 3.2 AT fuses (slow, ¼” x 1¼ ”, UL/CSA).
For 220 V (AC) ± 10%, use two 1.6 AT-fuses (slow, 5 x 20 mm, IEC127).
4. Switch ‘on’ the DECADE by the mains switch on the rear panel.
HPLC connections
Use proper eye and skin protection when working with solvents.
The installation of the flow cell, column and the injection valve (option) and
pulse damper (option) is schematically shown in Fig. 2.
5. If ordered, install the pulse damper. Connect the mounting plate with the
four bolts supplied to the pulse damper and mount the pulse damper with
four bolts. For detailed instructions on the pulse damper see page 75.
6. If ordered, the Antec supplied injection valve is equipped with a Vespel
rotor as a standard. For use of strongly alkaline solutions (pH > 10), this
should be replaced by a Tefzel rotor. Consult your supplier for details.
7. For installation of a manual injector remove the handle assembly and
place the body in the DECADE using the 2 factory-supplied mounting
screws. Re-assemble the handle assembly using the 2 set screws.