CHAPTER 3 The DECADE controller
23
Cycle
PULS SET2, PULS STAT
status
Displays the total duration of one pulse (t1 + t2 + t3). As from each cycle only
one data point is sent to the recorder output, this duration determines the
frequency of the recorder output (see page 51).
Cycle
SCAN SET
control
Controls the nature of the cycle: half, full and continuous. ‘Half’ means that
the cell potential runs from E1 to E2 and stops at E2. ‘Full’ means that the cell
potential runs from E1 to E2, and back to E1, and then stops. ‘Cont’ means
that the cell potential runs from E1 to E2 and back to E1 continuously (see
page 59).
Cycles
AUTO SET
control (master)
Controls the number of times a time file has to be repeated. If the DECADE is
master this number can be 1 - 99 or continuous. If the DECADE is slave the
number of cycles is 1 (see page 37).
CycleTime
AUTO SET
control
Controls the duration of a time file (max. 99’ 59”). When this time is reached
the execution of the time file stops. If programmed, the next run is started
(see page 37).
Decade
AUTO SET
control
Controls the status of the system, toggles between ‘Master’ and ‘Slave’. If
‘Master’ other parts of the HPLC system can be controlled by the DECADE
(see page 37).
DEL
(AUTO) PRGR
function
Deletes the current data line from the time file. Confirmation is asked for.
Time 00’ 00” can not be deleted (see page 37).
Drift
AUTO SET
control
Controls the driftmonitor in the auto mode, which intermits the execution of a