CHAPTER 10 Rear panel
71
C H A P T E R 1 0
Rear panel
TTL inputs
The inputs on the lower I/O connector strip, ‘start’, ‘reset’ and ‘auto zero’
require a TTL-low pulse (minimal 100 ms) to be activated. One TTL-low pulse
produces one activation, if more activations are required the next pulse must
be given after 100 ms TTL-high. If the input is kept low, only one activation will
result. See page 48 for detailed information on the I/O contacts.
RS232C
The RS232 option consists of a computer program, DECADE Dialogue for
Windows, and another DECADE EPROM with R-extension.
DECADE Dialogue for Windows is a multi-functional control bar which
handles all important settings of the DECADE via the RS232C port. The
program runs simultaneously with your favourite Windows-compatible
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com