CHAPTER 3 The DECADE controller
25
Interrupts the execution of a time file or scan. Toggles between 1 and 0.
Pressing the hold button again continues the time file or scan were it was
stopped (see page 37).
Icell
DC STAT, PULS STAT, SCAN STAT
status
Displays the true, non-compensated cell current, i.e. unaffected by zero or
offset. The cell current is one of the most important parameters in ECD.
INJ=I/L
DC STAT, PULS STAT
function/statu
s
Displays or switches the position of the injection valve, toggles between inject
(I) and load (L). If a manual injector with position sensor is applied, it echoes
the position of the injector. If an electrically actuated injector is used (optional)
it is possible to switch the injector with this function button.
MAIN
RANGE screen
function
Returns to the ‘MAIN’ screen.
MARK
DC STAT, PULS STAT
function
Triggers a marker signal on the recorder output.
NEXT
RANGE screen
function
Enters the next (xx SET) screen
Offs.
DC SET, DC STAT, (AUTO) PRGR, PULS SET1, PULS
STAT, SCAN SET, SCAN STAT
control
Controls the percentage offset, can be set between -50 and 50%. The offset
% on
both
outputs relates to the recorder range setting (see page 29).
Offs.
AUTO RUN
status
Displays the percentage offset during execution of a time file.
Outp.
(AUTO) PRGR
control
Controls the four output functions in the auto mode. The notation of the output