37XXXC OM
A-183
ALPHABETICAL LISTING
S
MENU
DESCRIPTION
GPIB COMMAND
SET SCALING
OR PRESS
<AUTOSCALE>
None
–LOG MAG–
None
RESOLUTION
XX.XXX dB/DIV
Sets the resolution for the vertical axis of the active chan-
nel’s displayed graph. Resolution can by set incrementally
using the keypad or rotary knob.
SCL; SCL?;
SCL2; SCL2?
REFERENCE VALUE
XXX.XXX dB
Sets the value at the reference line for the active channel
amplitude measurement on the log-magnitude graph. The
value can be set in increments of 0.001 dB using the key-
pad or rotary knob.
OFF; OFF?
OFF2; OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel’s amplitude
measurement on the log-magnitude graph. This is the line
about which the amplitude expands with different resolution
values. The reference line can be set to any vertical divi-
sion using the rotary knob.
REF; REF?;
REF2; REF2?
–PHASE–
None
RESOLUTION
XX.XX
°
/DIV
Sets the resolution for the vertical axis of the active chan-
nel’s displayed phase graph. Resolution can by set
incrementally using
the keypad or rotary knob.
SCL; SCL?;
SCL2; SCL2?
REFERENCE VALUE
XXX.XX
°
Sets the value at the reference line for the active channel
amplitude measurement on the phase graph. The value
can be set in increments of 0.001 dB using the keypad or
rotary knob.
OFF; OFF?
OFF2; OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel’s phase mea-
surement on the phase graph. This is the line about which
the phase expands with different resolution values. The ref-
erence line can be set to any vertical division using the ro-
tary knob.
REF; REF?;
REF2; REF2?
PHASE SHIFT
X.XX
°
Sets the value by which the active channel’s phase mea-
surement is shifted on the phase graph. The shift can be
set in increments of 0.01 degrees using the keypad or ro-
tary knob. This is useful when phase data is near the 180
degree rollover value.
PHO; PHO?
PRESS <ENTER>
TO RESUME CAL
On the CAL_SS1 menu, pressing the Enter key returns you
to the calibration setup or sequence.
None
Menu SS1 or CAL_SS1, Set Scaling 1
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...