The use of cables and/or adapters does not effect the final measure-
ment result, if they were in place for the calibration process. The vec-
tor error corrections established during the calibration process elimi-
nates cable and/or adapter effects as long as the ports used are stable
and exhibit good repeatability, which is the case if good quality compo-
nents are used. Figure 7-2 shows such a configuration.
Many calibration kits include adapters that are designed to have equal
phase length. These parts are called phase equal adapters (PEA).
Anritsu designs in-series adapters (e.g., K Connector M-M, M-F, F-F)
to be phase insertable when technically possible. When available, it is
good practice to use PEAs to establish test ports (Figure 7-3).
DISCUSSION
MEASUREMENT CALIBRATION
7-4
37XXXC OM
PORT 1 PORT 2
ADAPTERS
Figure 7-2. Using Adapters on the Test Port
33KKF50
33KFKF50
TEST SET CONNECTOR
KM
TEST PORT 1
KM
TEST PORT 2
KF
CABLE CONNECTOR
Figure 7-3.
Use of PEAs to Establish Test Ports
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...