10-3
CALIBRATIONS
Four types of calibration can be performed using AutoCal:
One-Port: S11 1-Port and S22 1-Port are 1-port calibrations per-
formed on the indicated port of the VNA and are equivalent to the tra-
ditional Open-Short-Load calibrations.
Full 2-Port: This type is equivalent to the traditional
Open-Short-Load-Thru (OSLT) calibration.
Thru Update: This type is a new form of calibration which is used to
update an existing 12-term calibration in the VNA. This calibration
could have been performed using any method of calibration which
yields 12 terms (LRL, LRM, AutoCal, or OSLT). Due to cable move-
ment and aging, the calibration may have degraded over time. The
Thru Update refreshes the calibration by measuring a Thru connec-
tion and updating the Transmission Tracking and Load Match calibra-
tion coefficients.
Adapter Removal: This calibration measures the characteristics of
male-male or female-female test port cables for subsequent measure-
ment of non-insertable devices. An adapter is required for this calibra-
tion. Adapter Removal requires two calibration procedures in order to
calculate the parameters and electrical length of the adapter.
10-4
DEFINITION
The following terms are used in explaining the calibration procedure
using the AutoCal module:
Thru: A thru is a connection of the two test ports. Two kinds of thru
connections are defined for the AutoCal calibration: (1) a Calibrator
thru is an internal path through the calibrator. (2) a True thru is a di-
rect cable connection between the test ports, with no intervening con-
nectors. The calibrator thru is not as accurate as a true thru, so the
you have the option during a calibration to use the more accurate
method, if necessary.
Switch Averaging: The mechanical AutoCal module uses electrome-
chanical switches to select calibration standards. These switches have
a small amount of non-repeatability (typically less than –55 dB). For
most calibrations, this is more than adequate because it is below con-
nector repeatability error. If desired, you can choose to reduce the ef-
fect of this non-repeatability in the mechanical module by using
Switch Averaging, which causes additional calibration measurements.
By setting a Switch Averaging factor larger than 1, switch repeatabil-
ity error will be reduced. The tradeoff is that calibration time will be
proportionally increased.
Isolation: For certain measurements which require accurate S
21
or
S
12
readings for very small values of those parameters, an isolation
CALIBRATIONS
AUTOCAL
10-4
37XXXC OM
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...