There are two ways of accomplishing this. The most obvious way is to
insert a length of line into the reference signal path to make both
paths of equal length (Figure 3-11, below). With perfect transmission
lines and a perfect splitter, we would then measure a constant phase
as we change the frequency. The problem using this approach is that
we must change the line length with each measurement setup.
Another approach is to handle the path length difference in software.
Figure 3-12 (left) displays the phase-vs.-frequency of a device. This de-
vice has different effects on the output phase at different frequencies.
Because of these differences, we do not have a perfectly linear phase
response. We can easily detect this phase deviation by compensating
for the linear phase. The size of the phase difference increases linearly
with frequency so we can modify the phase display to eliminate this
delay.
The 37XXXC offers automatic reference delay compensation with the
push of a button. Figure 3-13 (left) shows the resultant measurement
when we compensate path length. In a system application you can
usually correct for length differences; however, the residual phase
characteristics are critical.
NETWORK ANALYZERS
NETWORK ANALYZERS, A PRIMER
3-8
37XXXC OM
PHASE
DETECTOR
REFERENCE
SIGNAL
SPLITTER
TEST
SIGNAL
MICROWAVE
SOURCE
BOTH LINE
LENGTHS
NOW EQUAL
Figure 3-11.
Split signal where paths are of equal length
+180
+90
0
-90
-180
1.1 1.2 1.3 1.4
FREQUENCY,
GHz
M
E
AS
UR
E
D
P
H
A
S
E
SUBTRACT LINEAR
PHASE FROM
MEASURED PHASE
Figure 3-12.
Phase difference
increases linearly with frequency
0
1.1
1.2
1.3 1.4
FREQUENCY,
GHz
+2
+1
-1
-2
R
E
SU
L
T
A
N
T
PH
A
S
E
Figure 3-13.
Resultant phase
with path length compensation in
place
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...