Step 3.
The selections in the next menu to appear, MENU
ACAL, will depend on current instument conditions,
as follows:
a. THRU UPDATE lets you update the Thru cali-
bration of an active 12-term Calibration. This
updates the transmission frequency response
and load match coefficients.
b. START AUTOCAL lets you start a calibration
using the current setup.
c.
CHANGE AUTOCAL SETUP lets you set up a
new calibration, which is what we will do for
this example. This example also assumes that
you have selected the transmission medium
and, if waveguide, identified the cutoff fre-
quency.
USING AUTOCAL
AUTOCAL
10-10
37XXXC OM
MENU ACAL
AUTOCAL
AUTOCAL TYPE:
XXXXXXXX
CHANGE
AUTOCAL SETUP
START AUTOCAL
THRU UPDATE
CONNECT THROUGH
LINE BETWEEN
PORTS 1 AND 2
NUMBER OF AVGS
XXX
START THRU UPDATE
USE PREVIOUS
AUTOCAL SETUP
PRESS <ENTER>
TO SELECT
OR SWITCH
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...