Step 7.
Menu C1 (left), which appears next, lets you select
the number of frequency points at which calibration
data is to be taken. Of these choices, which were de-
scribed in paragraph 7-4, choose NORMAL (1601
POINTS MAXIMUM) for this example.
Step 8.
The next menu, C2 (below), lets you set your start
and stop frequencies. For this example, move cursor
to START, press
40
on keypad, and hit
MHz
termi-
nator key. Perform like operations for the STOP
choice, except make entry read 20 GHz. After setting
the frequencies, select NEXT CAL STEP and press
the
Enter
key.
Step 9.
When menu C3B (bottom left) appears, if you want
to change any of the parameters shown in blue let-
ters, place the cursor on that parameter and press
the
Enter
key. (These choices operate the same as
was described for menu C3 in paragraph 7-4.) For
this example, we change the waveguide parameters.
Move the cursor to WAVEGUIDE PARAMETERS
and press the
Enter
key.
OFFSET-SHORT CALIBRATION
MEASUREMENT CALIBRATION
7-30
37XXXC OM
MENU C1
SELECT
CALIBRATION
DATA POINTS
NORMAL
(1601 POINTS
MAXIMUM)
C.W.
(1 POINT)
N-DISCRETE
FREQUENCIES
(2 TO 1601
POINTS)
TIME DOMAIN
(HARMONIC)
PRESS <ENTER>
TO SELECT
MENU C2
FREQ RANGE OF
CALIBRATION
START
0.0400000000GHz
STOP
20.000000000 GHz
201 DATA PTS
0.099800000 GHz
STEPSIZE
MAXIMUM NUMBER
OF DATA POINTS
1601 MAX PTS
801 MAX PTS
401 MAX PTS
201 MAX PTS
101 MAX PTS
51 MAX PTS
NEXT CAL STEP
PRESS <ENTER>
TO SELECT
MENU C3B
CONFIRM
CALIBRATION
PARAMETERS
LOAD TYPE
BROADBAND
THROUGH LINE
PARAMETERS
WAVEGUIDE
PARAMETERS
XXXXXXXX
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...