Step 3.
When menu C11A (left) appears, move cursor to the
following:
a. STANDARD, then press
Enter
key. This selects
Standard (OSL) as the calibration method.
b. COAXIAL, then press
Enter
key. This selects
coaxial transmission line media.
c.
NEXT CAL STEP, then press
Enter
key. This
causes menu C11 to return to the screen.
Step 4.
When menu C11 reappears, confirm that the STAN-
DARD calibration method and COAXIAL line type
have been selected. Select NEXT CAL STEP and
press
the Enter
key to proceed. This brings up menu
C5.
Step 5.
Menu C5 (left) lets you select the type of calibration.
For this example, move the cursor to FULL
12-TERM and press
the Enter
key. This selection cal-
ibrates for all twelve error terms.
Step 6.
The next menu, C5D, lets you choose whether to in-
clude or exclude the error terms associated with
leakage between measurement channels. For a
normal calibration, you would choose to include
these error terms. Therefore, move the cursor to IN-
CLUDE ISOLATION ( STANDARD ) and press
the
Enter
key.
STANDARD (OSL) CALIBRATION
MEASUREMENT CALIBRATION
7-20
37XXXC OM
MENU C11A
CHANGE
CAL METHOD
AND LINE TYPE
NEXT CAL STEP
CAL METHOD
STANDARD
(NOT USED FOR
WAVEGUIDE)
OFFSET SHORT
LRL/LRM
TRANSMISSION
LINE TYPE
COAXIAL
WAVE GUIDE
MICROSTRIP
PRESS <ENTER>
TO SELECT
a.
b.
c.
MENU C5
CALIBRATION
TYPE
FULL 12-TERM
1 PATH
2 PORT
TRANSMISSION
FREQUENCY
RESPONSE
REFLECTION
ONLY
PRESS <ENTER>
TO SELECT
MENU C5D
SELECT USE
OF ISOLATION
IN CALIBRATION
INCLUDE
ISOLATION
(STANDARD)
EXCLUDE
ISOLATION
PRESS <ENTER>
TO SELECT
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...