Step 9.
When menu C3 (left) appears, if you want to change
any of the parameters shown in blue letters, place
the cursor on that parameter and press the
Enter
key. For this example, we will change them all, start-
ing with the top one. Move the cursor to PORT 1
CONN and press the
Enter
key.
Step 10.
In menu C4 (below), which appears next, move the
cursor to K CONN (M) and then press the
Enter
key.
This choice presumes that you have a K-Female con-
nector on the device-under-test (DUT). Remember, in
this menu you choose the connector type on the test
port, or the connector type that mates with the DUT
connector. When menu C3 returns, observe that K
CONN (M) is now shown in blue for the PORT 1
CONN choice.
STANDARD (OSL) CALIBRATION
MEASUREMENT CALIBRATION
7-22
37XXXC OM
MENU C3
CONFIRM
CALIBRATION
PARAMETERS
PORT 1 CONN
K CONN (M)
PORT 2 CONN
SMA (M)
REFLECTION
PAIRING
MIXED
LOAD TYPE
SLIDING
THROUGHLINE
PARAMETERS
REFERENCE
IMPEDANCE
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU C4
SELECT PORT 1
CONNECTOR TYPE
SMA (M)
SMA (F)
K-CONN (M)
K-CONN (F)
TYPE N (M)
TYPE N (F)
GPC-3.5 (M)
GPC-3.5 (F)
GPC-7
USER DEFINED
MORE
PRESS <ENTER>
TO SELECT
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...