37XXXC OM
A-23
ALPHABETICAL LISTING
C
MENU
DESCRIPTION
GPIB COMMAND
SELECT
CALIBRATION
DATA POINTS
None
NORMAL
(1601 POINTS
MAXIMUM)
Selects the standard calibration from a start to a stop fre-
quency that provides for up to 1601 equally spaced (except
the last) points of data for the defined frequency range.
NOC
C.W.
(1 POINT)
Selects the single frequency (C.W.) calibration sequence
that provides for 1 data point at a selected frequency.
CWC
N-DISCRETE
FREQUENCIES
(2 TO 1601
POINTS)
Selects the discrete frequency calibration mode that lets
you input a list of 2 to 1601 individual data point frequen-
cies.
DFC
TIME DOMAIN
(HARMONIC)
Selects the calibration mode for low-pass time-domain pro-
cessing.
TDC
PRESS <ENTER>
TO SELECT
Pressing the ENTER key implements your selection.
None
Menu C1, Select Calibration Data Points
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...