AD-4401A
CALF-
Item
Setting Value
Default
10
Tare and zero setting at unstable weight
0: Disable
1: Enable
1
11
Tare at negative gross value
0: Disable
1: Enable
1
12
Standard serial output, when the weight value
is overloaded or unstable.
0: Disable
1: Enable
1
13
BCD, RS-422/485, RS-232C output, when the
weight value is overloaded or unstable.
0: Disable
1: Enable
1
14 Weighing
mode
1: Simple comparison
normal batching
2: Simple comparison loss-
in-weight batching
3: Normal batching
4: Loss-in-weight batching
5: Check weighing 1
6: Check weighing 2
7: Check weighing 3
8: Check weighing 4
3
15
Load cell signal voltage at zero calibration
0.000000 to 2.200000
0.000000
16
(Load cell signal voltage at span calibration
CALF-15)
CALF-17 / Calibration weight value
at span calibration
0.000000 to 3.200000
2.000000
17
Maximum capacity at span calibration
0 to 999999
20000
18
Symbol used for header 2 of standard serial
output
1: GS/NT/TR
2: G/N/T
1
19
Symbol used for header 2 of RS-232C or
RS-422/485
1: GS/NT/TR
2: G/N/T
1
20
Automatic zero setting upon power-on
Zero setting range is ±10% of the maximum
capacity.
0: Disable
1: Enable
0
21 Zero
clear
0: Disable
1: Enable
1