14.9. RS-422/485, RS-232C Functions
For the function setting procedure, refer to “19. GENERAL FUNCTION SETTING.”
rS F-
Item
Setting Value
Default
01
Output weight value
1: Displayed weight value
2: Gross value
3: Net value
4: Tare value
5: Gross value
+
Net value
+
Tare value
6: Total weight
7: Number of accumulations
8: Total weight and number of accumulations
1
02
Data output timing
1: Interval output at the display update rate (FncF-03)
2: In synchronization with automatic print
3: In synchronization with manual print
4: Response to command
5: Interval output of comparison result and gross value at
100 times/s
6: Interval output of comparison result and net value at 100
times/s.
7: In synchronization with automatic/manual accumulation
8: Interval output at 100 times/s
1
03 Baud
rate
1: 600 bps
2: 1200 bps
3: 2400 bps
4: 4800 bps
5: 9600 bps
6: 19200 bps
7: 38400 bps
8: 115200 bps
5
04 Parity
0: None
1: Odd
2: Even
2
05 Character
bit
length
7: 7 bits
8: 8 bits
7
06
Stop bit length
1: 1 bit
2: 2 bits
1
07 Terminator
1:CR
2:CR LF
2
08
RS-422/485
electrical
specifications
1: RS-422 (rS F-09=0)
1: 4-wire RS-485 (rS F-09
0)
2: 2-wire RS-485
1
09
RS-422/485
slave address
0: None
1 to 99
0
10
Command response
delay time
0.00 to 3.00 s
0.06