AD-4401A Page
66
18.1. Calibration Functions
CALF-
Item
Setting Value
Default
01 Unit
0: None
1: g
2: kg
3: t
4: lb
2
(4)
02
Decimal point position
0: 0 (No decimal point)
1: 0.0
2: 0.00
3: 0.000
4: 0.0000
0
03
Minimum division
1, 2, 5, 10, 20, 50 d
1
04
Maximum capacity
0 to 999999
20000
05
Zero setting range
Set the zero setting range by the absolute value
in percent of the maximum capacity.
E.g.: Maximum capacity: 5000,
Zero setting range: 2
5000 x ±2% = ±100
Therefore, the zero setting range is within
±100.
0 to 100%
2
06 Zero
tracking
time
0.0: Not used
0.1 to 5.0 s
0.0
07 Zero
tracking
width
0: Not used
1: 0.5 d
2: 1.0 d
3: 1.5 d
4: 2.0 d
5: 2.5 d
6: 3.0 d
7:3.5 d
8:4.0 d
9:4.5 d
0
08
Stability detection time
0.0: Not used
0.1 to 5.0 s
1.0
09
Stability detection width
0: Not used
1 to 9 d
2