13.3. Output Functions
Select a data output timing (bCdF-02) from the following:
-
Interval output at the display update rate (FncF-03)
-
In synchronization with automatic print
-
In synchronization with manual print
Timing chart
The valid data output width is twice the strobe output width. Read the data within the valid data output width.
The strobe output width depends on the display update rate (FncF-03).
Display update rate
Strobe output width
20 times/s
20 ms
10 times/s
45 ms
5 times/s
95 ms
Update will stop within 50 ms after the hold input turns ON. Then, update will restart within 50 ms after the
hold input turns OFF.
13.4. BCD Output Functions
For the function setting procedure, refer to “19. GENERAL FUNCTION SETTING.”
bCdF-
Item
Setting Value
Default
01
Output weight value
1: Displayed weight value
2: Gross value
3: Net value
4: Tare value
1
02
Data output timing
1: Interval output at the display update rate
(FncF-03)
2: In synchronization with automatic print
3: In synchronization with manual print
1
03
Data output logic
1: Positive logic (Open-circuit is ON)
2: Negative logic (Short-circuit is ON)
2
04
Strobe output logic
1: Positive logic (Open-circuit is ON)
2: Negative logic (Short-circuit is ON)
2
Data output
Strobe output
Strobe output width
Valid data output width
Hold input
50 ms
50 ms