Page 17
AD-4401A
8. BATCH WEIGHING FUNCTIONS
Batch weighing is a procedure to automatically weigh the final weight.
Select a weighing mode (CALF-14) from the following:
-
Normal batching
-
Loss-in-weight batching
-
Simple comparison normal batching
-
Simple comparison loss-in-weight batching
8.1. Normal Batching
●
Weighing start
Operation: Function key (FncF-02), Control input
Setpoint values: Final weight, Free fall, Preliminary, Optional preliminary, Overweight, Underweight, Near-zero
Feeding process
1. The weighing start input is detected.
2. When the weighing start input delay time (Sq F-09) has passed, the large flow, medium flow and small
flow outputs turn ON.
3. When the large flow comparison disable time (Sq F-10) has passed and the weight value has exceeded
the optional preliminary value, the large flow output turns OFF.
4. When the medium flow comparison disable time (Sq F-11) has passed and the weight value has
exceeded the preliminary value, the medium flow output turns OFF.
5. When the small flow comparison disable time (Sq F-12) has passed and the weight value has exceeded
the free fall value, the small flow output turns OFF.
6. When the judging delay time (Sq F-13) has passed and the weight value is stable (Sq F-07), the weight
value is judged.
End of the feeding process
7. The weighing end output turns ON and the OK/Over/Under output of the judgment result turns ON.
Setting: OK/Over/Under output timing (Sq F-06)
8. When the weighing end output time (Sq F-14) has passed, the weighing end output turns OFF and the
OK/Over/Under output of the judgment result turns OFF.
●
Discharge start
Operation: Function key (FncF-02), Control input
Discharging process
1. The discharge start input is detected.
2. When the discharge start input delay time (Sq F-18) has passed, the discharge output turns ON.
3. When the weight value is below the near-zero value and the discharge end delay time (Sq F-19) has
passed, the discharge output turns OFF.
●
Near-zero
When the weight value is compared with the setpoint and the condition is met, the control output turns ON.
Condition
Control output
Weight value
Near-zero (Sq F-26)
Near-zero