PROCESS DESCRIPTION
P
ROCESS
M
ANUAL
4.3-5
4.3.7
Recipe example: LPCVD silicondioxide from LTO SiH
4
and O
2
1
00 START RECIPE
2
Message LOAD/UNLOAD [4]. Sonalert alarm No
3
Normal recipe 00 Zone1 425.0 °C Slope 10.00 °C/min Zone2 425.0 °C Slope 10.00
°C/min Zone3 425.0 °C Slope 10.00 °C/min , profile table A
4
Boat to 2000.0 mm at 300.0 mm/min with oscillation speed of 0.0 mm/min.
Variable Command: No
5
Gas N2 [1] at 0.00 [SLM] Variable Command: No
6
Gas O2 [2] at 0 [SCCM] Variable Command: No
7
Gas SIH4 [3] at 0.0 [SCCM] Variable Command: No
8
Gas PRESSURE [8] at 0 [MTOR] Variable Command: No
9
Digital out N2 PURGE[9] = OFF,N2PROCES[10] = OFF,PROCESS[11] =
OFF,EVACSIH4[12] = OFF,,,SOFTSTRT[15] = OFF,MAINVAC[16] = OFF
Variable Command: No
10
Alarm Limit Setting for All Analog Output Channel at 0 %
11
Alarm on digital in PRESS N2[1] = ON,PRESSAIR[2] = ON,doorclsd[3] =
OFF,VACFAIL[4] = OFF,,,TEMP SCR[7] = ON,EXCESS[8] = ON
12
Alarm on digital in FANCONTR[9] = ON,,,OPERATE[12] = ON,WARNING[13]
= ON,FAILURE[14] = ON,,
13
01 BOAT OUT
14
Message BOAT OUT [3]. Sonalert alarm No
15
Time: 000:10:00 (hr:min:sec) Variable Command: No
16
Boat to 10.0 mm at 300.0 mm/min with oscillation speed of 0.0 mm/min. Variable
Command: No
17
Gas N2 [1] at 10.00 [SLM] Variable Command: No
18
Digital out N2 PURGE[9] = ON,N2PROCES[10] = ON,PROCESS[11] =
OFF,EVACSIH4[12] = OFF,,,SOFTSTRT[15] = OFF,MAINVAC[16] = OFF
Variable Command: No
19
Wait for boat to reach setpoint
20
02 LOAD WAFERS
21
Message LOADWAFERS [5]. Sonalert alarm Yes
22
Wait for operator
23
03 BOAT IN
24
Message BOAT IN [2]. Sonalert alarm No
25
Time: 000:10:00 (hr:min:sec) Variable Command: No
26
Boat to 2000.0 mm at 300.0 mm/min with oscillation speed of 0.0 mm/min.
Variable Command: No
27
Wait for boat to reach setpoint
28
Wait on digital in PRESS N2[1] = OFF,PRESSAIR[2] = OFF,doorclsd[3] =
ON,VACFAIL[4] = OFF,,,TEMP SCR[7] = OFF,EXCESS[8] = OFF
29
04 EVAC 1
30
Message EVACUATE 1 [8]. Sonalert alarm No
31
Time: 000:01:00 (hr:min:sec) Variable Command: No
32
Gas N2 [1] at 0.00 [SLM] Variable Command: No
33
Digital out N2 PURGE[9] = ON,N2PROCES[10] = ON,PROCESS[11] =
OFF,EVACSIH4[12] = OFF,,,SOFTSTRT[15] = ON,MAINVAC[16] = OFF
Variable Command: No
34
05 EVAC 2
35
Message EVACUATE 2 [9]. Sonalert alarm No
36
Time: 000:05:00 (hr:min:sec) Variable Command: No