Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The equipment or system is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment or system should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment – Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV for power supply lines
±0.250 kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips, Short
Interruptions and Voltage
Variations on Power Supply
Input Lines IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of the equipment
or system requires continued
operation during power mains
interruptions, it is recommended
that the equipment or system be
powered from an uninterruptible
power supply or a battery.
Power Frequency (50/60 Hz)
Magnetic Field IEC 61000-4-8
3 A / m
Not Applicable
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or
hospital environment.
Electromagnetic Compliance Data for iCE 30m Series
NOTE:
UT is the A.C. mains voltage prior to application of the test level.