Amico Clinical Solutions Corp.
!
POSSIBLE EQUIPMENT DAMAGE
• Cleaning and disinfecting agents used on this lighting system must be certified by their manufacturer to be
compatible with polycarbonate.
• Use only recommended cleaning/disinfecting and/or anti-static agents on this light. Use of alcohol or aerosol
spray cleaner/disinfectants containing a substantial amount of alcohol in the formula can damage the
polycarbonate lens. Use manufacturer recommended cleaners only.
• Use of any disinfectant solutions OTHER than Germicidal Surface Wipes and Disinfecting/Deodorizing/Cleaning
Wipes may cause discoloration or deformation of the polycarbonate lens surface. Other solutions have NOT
been tested for compatibility or effectiveness.
• Cleaners intended for use on floors must NOT be used to clean this equipment.
• Prevent leakage of fluids into interior lighthead. Do not scratch optical coating on accessible portions of optic
assembly when cleaning: always wear rubber gloves and use only a clean, white, lint-free cloth when wiping
surfaces.
• Do not touch the glass portion of the lamp with bare fingers. Skin oils can cause deterioration of material
leading to possible failure of the lamp.
• Do not bump lightheads into walls or other equipment.
• During all mounting steps, verify that wires are not pinched between spring arms or lightheads.
• Avoid cross-threading brake screws, align them with extreme care.
• Accessories or replacement parts not purchased from Amico are not to be used as they may negatively affect
the equipment or result in equipment damage.
• During the mounting of the LED lights, the entire system (including the ceiling attachment) must be
disconnected from mains. Dismounting of the lights from the spring arms or dismounting the sliding contacts
inside the arms is to be done ONLY AFTER DISCONNECTING THE ENTIRE SYSTEM FROM MAINS.
Otherwise the main control board will be damaged.
• Use only Amico supplied fuses. Any higher rating could cause damage to the LEDs and reduce their life.
Accessories:
• Cameras - Sony Models EH 6300, EH 6500, EV 7100 and EV 7500
• Handles - From Amico and other approved manufacturers
Safety Instructions