8
Amico Clinical Solutions Corp.
Technical Data for iCE 30m Lightheads
!
Never combine beams from more than two lampheads, as the resulting irradiance at the spot can exceed
1000 W/m² resulting in a higher than normal temperature.
Upon request, Amico Clinical will make available the circuit board diagrams, component parts list,
descriptions, calibration instructions or other information that will assist service personnel in conducting
repairs to the iCE 30m.
Recommended training for safe use of the iCE Light is approximately 30 minutes for users and two hours for
biomedical staff.
Performance
Units
160,000
Lighthead Diameter
in (cm)
30 (76)
Light Intensity (Central Luminance)
lx (fc)
160,000 (14,870)
Brightness Control
%
50 - 100 (5% in Endo Mode)
Light Field Diameter
in (cm)
7.6 - 12 (19 - 31)
1
Depth of Illumination (L1+L2) at 60%
in (cm)
16.8 (42.7)
Color Rendering Index (Ra)
97
2
R9
>90
R13
97
2
Color Temperature
K
3,500⁄4,000⁄4,500⁄ 5,000
3
Shadow Dilution
Single Mask
%
63
Double Mask
%
52
Cavity
%
99
Single Mask with Cavity
%
61
Double Mask with Cavity
%
50
Number of LEDs (main)
80
Number of Other LEDs
320
LED Service Life
hours
50,000
Total Radiant Power at Max Intensity
W/m
2
515
Power Consumption at Lighthead
w @ 24 VDC
50
Electricity (VAC)
Universal VAC
Emergency Bypass
Yes
Mounting Options
Ceiling⁄ Wall⁄Mobile Stand
Certificates
TUV
All tests conducted per 60601-2-41 standard • ¹LFD has a tolerance of 5% • 2CRI has a tolerance of 2.5% • 3CCT has a tolerance of 15%
Содержание ice 30m
Страница 1: ...Installation and Operation Instructions Surgical Lights w w w a m i c o c o m...
Страница 6: ...Typical Drawings Single Ceiling Mount Single Ceiling Mount for Low Ceiling Height Dual Ceiling Mount...
Страница 46: ...46 Amico Clinical Solutions Corp Operation The Amico iCE 30m LED Light Expanded Beam...