Troubleshooting | 5-7
To check if a
calibration gas
solenoid is stuck
open:
Verify that no drive signal is present and check for flow on the
RCU flow meter.
If flow is indicated when no solenoids are ener
-
gized, a solenoid is stuck open.
Shut off your calibration gases
(one at a time) until the flow drops to zero. This identifies the
defective solenoid valve (replace the solenoid valve, being sure
that the solenoid O-ring seals are properly positioned).
Calibration Gas
Time Inadequate
If you are having problems running an auto calibration, you
may not have allowed the calibration gases enough time to flow
through the sensor and stabilize. To correct this problem, select
Inject Cal Gas
from the
Diagnostic
menu.
Turn on each calibration gas and determine how long it takes
for each gas to stabilize on the control unit display. Then add a
one minute buffer to each of these times. Also make sure the
cylinder regulators are set to the correct pressure.
See the Flow
Section of the System Interconnect Drawing for the system flow
and pressure requirements
. This drawing is included with your
Installation package.
Select
Cal Gas Duration
from the
Calibration
menu to set cali-
bration gas times.
If performing a manual calibration, ensure that you are waiting for the
reading to stabilize on the display before switching to the next calibra-
tion gas.
AC Power Checks
Loss/Inadequate AC Voltage
to the Sensor
Measure the AC voltage to the sensor board at terminals L and
N. Ensure that this voltage is sufficient. Check the measure
-
ment technique used by the Volt meter (for example, RMS,
average, peak, etc.). Specifications are based on RMS mea
-
surements.
Furnace Checks
Open Furnace
Disconnect power to the control unit and the sensor. With Ohm
meter, measure across Terminals “FURN” on the sensor board.
The resistance of the furnace should be:
32 ohms (±10%)
If the furnace resistance is not within allowable tolerances,
replace the furnace.
Loss of AC Power
to the Furnace
Verify the correct line voltage at L and N of the sensor board.
With line voltage present, check the voltage at terminals FURN
on the sensor board.
If voltage is present
and the system is not
heating, remove power and check the furnace resistance (see
“Furnace Checks” section).
If no voltage is present
at terminals
FURN on the sensor board, replace the electronics.
NOTE
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...