Troubleshooting | 5-5
Thermocouple Checks
Open Thermocouple
Remove power to the sensor. Measure across terminals
“+TC-” on the sensor board with an ohm meter. If an open
is measured, replace the thermocouple.
Shorted/Failed Thermocouple
Check that the thermocouple leads are not shorted to
chassis ground by using an Ohm meter to measure
between terminal TC+ on the sensor board and chassis
ground, and between terminal TC- on the sensor board
and chassis ground. If shorted, replace the thermocouple.
To verify the operation of the thermocouple itself, remove
the thermocouple from the sensor and heat its ceramic tip
to a known temperature. Measure the millivolt output with
a proper temperature indicating meter (type K thermocou
-
ple). If the thermocouple reads inaccurately, replace it.
Reversed Thermocouple Wires
First, view the cell temperature via the Home screen or
Diagnostic screen. If the thermocouple leads are reversed,
the displayed temperature will be decreasing as the sensor
warms up (this will usually happen at start-up or after you
replace a thermocouple). This indicates that the thermo-
couple wiring is reversed. If you just replaced a thermo
-
couple, check the leads from the thermocouple to the
sensor board: yellow wire connects to terminal TC (+), red
wire connects to terminal TC (-).
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...