Installation Guide | 3-3
Mechanical Installation
This section describes how to mount and connect gases to your WDG-V
sensor. This includes the following:
• Sample Inlet Probe Installation (Wall/Flange Mount)
• Sample Inlet Probe Installation (Floor Mount)
• Exhaust Tube Installation (optional)
• Sensor Mounting (Wall/Flange Mount)
• Sensor Mounting (Floor Mount)
• Remote Calibration Unit Mounting and Plumbing (optional)
Sample Inlet Probe Installation (Wall/Flange Mount)
The WDG-V with Blow Back Option unit consists of a bolt-on attachment
to the standard WDG-V analyzer for use in high particulate applications.
Figure 3-1a and 3-1b. WDG-V with Blow Back and internal components.
Box Temperature is controlled to 150 ºC to keep above dew point.
Component
Function
Accumulator Ball
Warms instrument air to avoid cold air blow
backs and condensation.
RTD in Box
Alarms if temperature goes out of range.
Resettable Over-Temp Switch
Prevents damage to internal components.
1. Using a standard 3”-300# mounting flange (shown in Figure 3-1b)
mount the analyzer to the process stack.
Other options for mounting are available.
NOTE
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...