Specifications | 1-3
Sensor : WDG-VC
Output Range
From 0-1% to 0-100%
Oxygen
Combustibles
From 0-1% to 0-100%
From 0-2,000 ppm to 0-10,000 ppm or from 0-1% to 0-5%
Accuracy
Oxygen
± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles
± 2% of full scale output range
Response O2
90% of a process step change < 6 secs.
Drift
< 0.1% of cell output per month (< 0.005% O
2
per month with 2% O
2
applied)
Aspirator Air
Requirements
≈
3 scfh (1.4 L/min) at 80 to 100 psig (5.63 to 7.04 kg/cm
2
)
Max. Flue Gas Temp / Probe Type / Lengths
1300°F (704°C)
1875°F (1024°C)
3000°F (1649°C)
316 SS
310 SS
Hexoloy
®
36” - 108” (91 cm - 271 cm)
36” - 108” (91 cm - 271 cm)
24” - 72” (60 cm-182 cm)
Max. Sample
Dewpoint
392°F (200°C)
Sample Pressure
± 10 inch water gauge
Environment
Ambient Temp.:
-13°F to 149°F (-25°C to 65°C)
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity:
5% to 95%, non-condensing
Max. Altitude:
5000 meters
Enclosure
Power Requirements
115 VAC, ± 10%, 47-63 Hz, 950 VA max.
230 VAC, ± 10%, 47-63 Hz, 950 VA max.
Calibration Gas Requirements
Use calibration gases @ 10 PSIG, 3.0 SCFH (1.4 L/min.)
O
2
Span Gas:
Air or from 1.0 to 100% O
2
, balance N
2
O
2
Zero Gas &
Comb Zero Gas:
2% or from 0.1 to 10% O
2
, balance N
2
Comb Span Gas:
60 to 80% (ppm ranges) of the selected comb. recorder output range
in certified mixtures of
CO + H
2
, 3-4% O
2
, balance N
2
.
System Compliance
EMC Compliance:
Safety Compliance:
2004/108/EC
73/23/EEC
NOTES:
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas (without flame arrestors).
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...