2-10 | Thermox WDG-V / VC / VCM with Blow Back
Temperature Sensors
There are two tightly controlled temperature zones in the WDG-V. The
Oxygen Cell Temperature and the combustible block temperature (CM
units only). (On oxygen-only units the box temperature is controlled with
an RTD mounted on the aspirator block.)
Cell Temperature
The oxygen cell is controlled to 680 °C. The cell temperature is measured
by dual, Type K thermocouples. Dual thermocouples are used for self
check of the cell temperature. The thermocouples are collocated in a
single assembly. There is no distinction between the two thermocouples
as they are effectively at the same location in the assembly.
Flow Sensor Heater Element (Black Wire)
Flow Sensor Sense Element (Brown Wire)
Combustible/Box Temperature
The combustible block/ box temperature is controlled to 225 °C. This
temperature is measured by 100 Ohm platinum RTD. The RTD is embed-
ded in the combustible block. On oxygen only units, the RTD is mounted
on the aspirator block. All internal plumbing components are maintained
above 200 °C.
Blow Back Thermostat
Located in the blow back enclosure to control temperature to 150 °C.
Heaters
There are three main heaters in the WDG-V, the cell heater the sensor box
heater and the blow back box heater. In CM units, an additional heater is
used for fine temperature control of the combustible block.
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...