3-28 | Thermox WDG-V / VC / VCM with Blow Back
Notes About Ducting
Ducting Between Pressurized Enclosure and Inlet
The intake ducting to a compressor should not normally pass through a
hazardous area. If the compressor intake line passes through a hazardous
area, it should be constructed of non-combustible material and protected
against mechanical damage and corrosion. Adequate precautions should
be taken to ensure that the ducting is free from leaks in case the internal
pressure is below that of the external atmosphere (see Annex C). Addi-
tional protective measures, for example, combustible gas detectors, should
be considered to ensure that the ducting is free of flammable concentra-
tions of gas vapor.
Additional Purge Time to Account for Ducting
The purge duration should be increased by the time necessary to purge
the free volume of those associated ducts which are not part of the equip-
ment by at least five times their volume at the minimum flow rate speci-
fied by the manufacturer.
Power for Protective Gas Supply
The electrical power for the protective gas supply (blower, compressor,
etc.) should be either taken from a separate power source or taken from
the supply side of the electrical isolator for the pressurized enclosure.
Enclosure Maximum Overpressure
The user should limit the pressure as specified by the manufacturer.
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...