Installation Guide | 3-11
Calibration/Aspirator Air
Required Calibration Gases and Tubing
See Chapter 2 for information on calibration gases. The span gas is
the high gas; the zero gas the low gas.
The span gas must be 10 times greater than the zero gas. For example, if
the zero gas is 1% O
2
, the span gas must be 10% O
2
or higher.
O
2
Span Gas
Instrument air (20.9%) or from 1.0% to 100% O
2
.
O
2
Zero Gas
From .1 to 10% O
2
, balance N
2
Combustibles Span Gas
If using a WDG-VC analyzer, you must also have a combustibles
span gas as follows:
60 to 80% of the full combustibles operating range in certified
mixtures of CO, with an equal amount of O
2
, balance N
2
.
For example, if using an operating range of 0 to 2000 ppm combustibles,
use 1200 ppm CO, 1 to 2% O
2
, balance N
2
.
Methane Span Gas
• If using a WDG-VCM analyzer, you also must have a methane span
gas as follows:
- 2% CH
4
, 8 to 10% O
2
, balance N
2
.
ASPIRATOR AIR
DRAIN VALVE
INSTRUMENT AIR INLET
ELECTRONICS SECTION
C L
LOCATION
CL
[
[
[
[
]
]
]
]
Figure 3-12. Sensor Inlet Connections (Wall/Flange Mount).
NOTE
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...