PC Configurator Software | A-13
Calibration Button
Use the Calibration Menu to configure the sensor for calibration, perform
the calibration, and view the calibration history.
TABS
OPTIONS
CALIBRATION
Set gases to calibrate. Set gas values
and timers. Perform a manual calibra-
tion now.
CALIBRATION HISTORY
Plots the history of gases calibrated.
AUTO CALIBRATION
CONFIGURATION
Set time and frequency for Auto Cal.*
CALIBRATION GAS VALUES
Set values and timers for calibration
gases to perform a manual calibration at
a later time.
*
Automatic calibration requires a factory-provided remote calibration unit (RCU)
that automatically switches the calibration gases.
All measurements require a two-point calibration: span and zero, so
the zero will always be performed.
Calibrations are only allowed when the sensor is in Normal mode.
Check the Status bar at the bottom of the Home screen if you are
unsure.
All gases selected under Sensor Configuration (O
2
, combustibles, meth-
ane) require calibration but do not have to be calibrated at the same time.
However, calibration will always occur in the same sequence.
1. O
2
Span
2. Zero
3. Combustibles Span
4. Methane Span
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...